Adept Ultrasonic Flowmeter – UFM 6740
Battery Powered Water Meter
Introduction
Adept UFM 6740 inline type Flanged Battery Operated Ultrasonic Flowmeter provides abundant capabilities for accurate liquid flow measurement. It is widely used in remote location applications where Field Electric supply is not available.
Salient Features








Applications




Product Selection Guide
Parameters | Specifications |
---|---|
Technical characteristics | Low power consumption, digital process technology with stable piezoelectric sensor |
Accuracy | ±1% of reading (Optional ±0.5%) |
Repeatability | 0.2% of Span |
Display | Total flow, flow rate, velocity, working time & date, signal intensity, quality & battery status |
Output | RS 485, 4 – 20 mA, Pulse for total flow (All outputs need external power supply 24 V DC) |
Communication interface | GSM, RS 485 (Modbus RTU), UART |
Operation | Inside keypad available |
Sampling period | 0.5 sec |
Environment temp. | Up to 65°C |
Water temp. | 5 to 70°C |
Power supply | Li-ion battery, solar powered |
Enclosure class | IP 68 |
Pipe size | DN 40 – 600 |
Flange standard | ANSI, IS, DIN PN, BS |
Pressure transmitter | Range 0-16 kg/cm², IP 68, 1/2″ male threaded, accuracy 0.25% |
Advanced Technology Flow Instrumentation Solutions
Adept is one of the largest manufacturers of Electromagnetic and Ultrasonic Flowmeters in India and has been acquired by Yokogawa Electric Corporation. Yokogawa provides advanced solutions in the areas of measurement, control, and information to customers across a broad range of industries, including energy, chemicals, materials, pharmaceuticals and food. The acquisition will help to leverage the strong presence of Adept in the Indian market with the technological prowess of Yokogawa.
Adept range of Flowmeters is characterised by advanced technology, excellent quality and high reliability to ensure accurate and precise measurement of fluid flow. We have a large manufacturing facility with two factories operational in Pune, India, certified with IMS (Integrated Management System). This is backed by our in-house Design & Development Centre and the Calibration Laboratory certified with ISO/IEC 17025:2017. More than 70,000 Flowmeters supplied by us are operational in over 25 countries across the world, meeting the stringent requirements of different sectors. The reliability of our products and the dependability of our services are evident from repeat orders from our customers. Over 40% of our business comes from existing customers.
Related products
Adept BTU Meter 6430 Measures Energy Flow precisely and conveniently
Measures Energy Flow precisely and conveniently
Introduction
Salient Features










Applications









Product Selection Guide
Parameters | Specifications |
---|
Power supply | 80 – 300 V AC |
Flowmeter input | 4 – 20 mA DC from Flowmeter (Ultrasonic/Electromagnetic/Any other) |
Temparature input | Matched Pair of PT-100 Sensors or PT 1000 sensors , Standard length – 6m |
Display | 2 Line LCD |
Communication | RS 485 (Modbus RTU), BACnet IP (Optional) |
Mounting | Wall Mount |
Accuracy | ± 0.1% of measured value |
Output – Analog | 4 – 20 mA DC |
Output – Contact | Relay (Optional) |
Energy function | Heat/Cool |
Energy display | W, J, BTU, Cal |
Sampling period | 1 sec. |
Environment temp. | Up to 65°C |
Water temp. | 0 to 50°C (Please consult factory for any other liquid.) |
Enclosure class | IP 65 |
FAQ
Adept Electromagnetic Flowmete – MagBP 6440
Introduction
Salient Features








Applications









Product Selection Guide
Parameters | Specifications |
---|---|
Nominal dia (mm) | 10 to 1000 |
Working pressure (kg/cm2) | 10, 16, 25, 40 |
Working temperature | Up to 55°C |
Electrode material | SS 316L Std.* |
Sensor lining | Std. Rubber* |
Display version | Integral/Remote |
Measuring tube material | SS 304 Std.* |
Sensor housing material | Std. CS* |
End connection | Flange/Wafer/Tri-clamp/SMS |
Flange standard | ANSI 150* |
Measuring range | 0.2 to 12 m/sec. Bidirectional |
Accuracy % of measured value | ±0.5% |
Repeatability | ±0.2% of Span |
Display | LCD |
Display units | All standard engineering units in m³, litre, gallon, ft³, imperial gallon |
Output | Pulse* |
Power supply | Battery powered (5 years’ battery life) |
Protection class for sensor | Std. IP 65 Option IP 67/IP 68 for flow tube in remote type |
Protection class for transmitter | IP 67 |
Cable length for remote | Std. 10 m* |
Installation | Inline flanged type |
Adept Electromagnetic Flowmete – MagFlow 6415
Introduction
Salient Features








Applications









Product Selection Guide
Parameters | MagFlow 6415 |
---|---|
Nominal dia (mm) | 10 to 3000 |
Working pressure (kg/cm2) | 10, 16, 25, 40 |
Working temperature | Integral PTFE – 120°C Remote PTFE – 180°C Others – 70°C |
Electrode material | SS 316L Std.* |
Sensor lining | Std. Rubber* |
Display version | Integral/Remote |
Measuring tube material | SS 304 Std.* |
Sensor housing material | Std. CS* |
End connection | Flange/Wafer/Tri-clamp/SMS |
Flange standard | ANSI 150* |
Measuring range | 0.2 to 12 m/sec. Bidirectional |
Accuracy % of measured value | ±O.5% (±O.2% consult factory) |
Repeatability | ±0.2% of Span |
Display | Graphic LCD |
Display units | All standard engineering units in m³, litre, gallon, ft³, imperial gallon |
Output | Std. 4 – 20 mA, Pulse, RS 485 |
Dual power supply | 12 – 60 V DC or 80 – 300 V AC/DC |
Protection class for sensor | Std. IP 67 Option IP 68 for flow tube in remote type |
Protection class for transmitter | IP 67 |
Cable length for remote | Std. 10 m* |
Grounding | Built-in electrode |
Installation | Inline |
Adept Electromagnetic Flowmete – EcoMag 6420
Introduction
Salient Features










Applications









Product Selection Guide
Parameters | Specifications |
---|---|
Nominal dia (mm) | 10 to 200 |
Working pressure (kg/cm2) | 5 |
Working temperature | Up to 55°C |
Electrode material | SS 316L Std.* |
Sensor lining | NA |
Display version | Integral/Remote |
Measuring tube material | HDPE |
Sensor housing material | HDPE |
End connection | Flange |
Flange standard | ANSI 150* |
Measuring range | 0.2 to 12 m/sec. Bidirectional |
Accuracy % of measured value | ±1% |
Repeatability | ±0.2% of Span |
Display | 2 line LCD |
Display units | All standard engineering units in m³, litre, gallon, ft³, imperial gallon |
Output | Std. 4 – 20 mA* |
Power supply | 12 – 60 V DC or 80 – 300 V AC/DC Solar powered |
Protection class for sensor | Std. IP 65 Option IP 67/IP 68 for flow tube in remote type |
Protection class for transmitter | IP 67 |
Cable length for remote | Std. 10 m* |
Installation | Inline flanged type |
Adept Electromagnetic Flowmeter – MagProbe 6450
Introduction
Salient Features








Applications




Product Selection Guide
Parameters | Specifications |
---|---|
Nominal dia (mm) | > 100 |
Working pressure (kg/cm2) | 20 |
Working temperature | Up to 120°C |
Electrode material | SS 316L Std.* |
Sensor lining | NA |
Display version | Integral/Remote |
Measuring tube material | SS 316 Std.* |
Sensor housing material | NA |
End connection | NA |
Flange standard | NA |
Measuring range | 0.2 to 12 m/sec. Bidirectional |
Accuracy % of measured value | ±2% |
Repeatability | ±0.2% of Span |
Display | 2 line LCD |
Display units | All standard engineering units in m³, litre, gallon, ft³, imperial gallon |
Output | Std. 4 – 20 mA* |
Power supply | 12 – 60 V DC or 80 – 300 V AC/DC Solar powered |
Protection class for sensor | Std. IP 68 |
Protection class for transmitter | IP 67 |
Cable length for remote | Std. 10 m* |
Installation | Insertion type with use of isolating ball valve assembly on pipeline |
Adept Electromagnetic Flowmeter MagFlow 6410
Introduction
Salient Features








Applications









Product Selection Guide
Parameters | Specifications |
---|---|
Nominal dia (mm) | 10 to 3000 |
Working pressure (kg/cm2) | 10, 16, 25, 40 |
Working temperature | Integral PTFE – 120°C Remote PTFE – 180°C Others – 70°C |
Electrode material | SS 316L Std.* |
Sensor lining | Std. Rubber* |
Display version | Integral/Remote |
Measuring tube material | SS 304 Std.* |
Sensor housing material | Std. CS* |
End connection | Flange/Wafer/Tri-clamp/SMS |
Flange standard | ANSI 150* |
Measuring range | 0.2 to 12 m/sec. Bidirectional |
Accuracy % of measured value | ±O.5% (±O.2% consult factory) |
Repeatability | ±0.2% of Span |
Display | 2 line LCD |
Display units | All standard engineering units in m3, litre, gallon, ft3, imperial gallon |
Output | Std. 4 – 20 mA* |
Power supply | 12 – 60 V DC or 80 – 300 V AC/DC Solar powered |
Protection class for sensor | Std. IP 65 Option IP 67/IP 68 for flow tube in remote type |
Protection class for transmitter | IP 67 |
Cable length for remote | Std. 10 m* |
Installation | Inline flanged type |
Adept IoT Gateway 5000
Introduction
Specifications








Adept Ultrasonic Flowmeter – UFM 6730
Introduction
Salient Features









Applications









Product Selection Guide
Parameters | Single Path | Two Path | Four Path |
---|
General | |||
No. of Sensors | 2 | 4 | 8 |
Operating principle | Transit time technology | Transit time technology | Transit time technology |
Operating pressure | 20 kg/cm² max | 20 kg/cm² max | 20 kg/cm² max |
Application | Raw, clear, sewage water & homogeneous liquids | Raw, clear, sewage water & homogeneous liquids | Raw, clear, sewage water & homogeneous liquids |
Transmitter | |||
Power supply | 24 V DC ± 20% 110 V AC ± 15% 240 V AC ± 15% Solar powered | 24 V DC ± 20% 110 V AC ± 15% 240 V AC ± 15% Solar powered | 24 V DC ± 20% 110 V AC ± 15% 240 V AC ± 15% Solar powered |
Velocity | ±12 m/sec. Bidirectional | ±12 m/sec. Bidirectional | ±12 m/sec. Bidirectional |
Display | 2 lines, 16 characters LCD with backlight flow rate, velocity, totaliser | 2 lines, 16 characters LCD with backlight flow rate, velocity, totaliser | 2 lines, 16 characters LCD with backlight flow rate, velocity, totaliser |
Units | User configurable English/Metric | User configurable English/Metric | User configurable English/Metric |
Totaliser | FWD, NET, REV | FWD, NET, REV | FWD, NET, REV |
Output – Analog | 4 – 20 mA Std.* | 4 – 20 mA Std.* | 4 – 20 mA Std.* |
Accuracy | ±1% of reading @ rate >0.5 m/sec. | ±0.5% of reading @ rate >0.5 m/sec. | ±0.5% of reading @ rate >0.5 m/sec. |
Repeatability | 0.2% of Span | 0.2% of Span | 0.2% of Span |
Protection class | IP 65 | IP 65 | IP 65 |
Dimensions | 251X190X96 mm | 500x400x300 mm | 500x400x300 mm |
Weight | < 2.5 kg | < 4.5 kg | < 6.5 kg |
Security | Keypad lockout, access code enabled | Keypad lockout, access code enabled | Keypad lockout, access code enabled |
Self diagnostic feature | Available | Available | Available |
Facility to measure and display each path velocity | NA | Available | Available |
Ability to work even though one or more path does not work | NA | Available | Available |
Flow and totaliser unit programmability | Available | Available | Provided |
Password protection for configuration data | Available | Available | Available |
Totaliser value retention in the event of power fail condition | Available | Available | Available |
RS 485 Modbus RTU Protocol | Available | Available | Available |
Lightening protection units | Available | Available | Available |
Sensor | |||
Suited temp. – Std. | -40°C to 120°C | -40°C to 120°C | 40°C to 120°C |
Pipe size | 50 – 6000 mm dia | 350 – 6000 mm dia | 1000 – 6000 mm dia |
Protection class | IP 65, IP 67, IP 68 | IP 65, IP 67, IP 68 | IP 65, IP 67, IP 68 |
Sensor dimensions | 186 mm with 20 mm bore size | 186 mm with 20 mm bore size | 186 mm with 20 mm bore size |
Weight | < 1.5 kg | < 1.5 kg | < 1.5 kg |
Cable length | 6 m std. can be extended up to 300 m | 6 m std. can be extended up to 300 m | 6 m std. can be extended up to 300 m |